Devices: |
|
Measurement mode |
Active and Passive Mode |
Measurement Range (BAC) |
(0-550)mg/100ml BAC |
Accuracy: |
|
Range (0-88)mg/100ml |
± 4.4 mg/100ml |
Range (88-220)mg/100ml |
± 5 % absolute |
Range >220 mg/100ml |
± 20 % absolute |
Measurement Range (BrAC) |
(0-2.5)mg/L BrAC |
Accuracy: |
|
Range (0.00-0.40) mg/L |
± 0.020 mg/L |
Range (0.40–1.00) mg/L |
± 5 % absolute |
Range >1.00 mg/L |
± 20 % absolute |
Condition: |
|
Operation Temperature Range: |
– 5 °C – +50 °C |
Storage Temperature Range: |
– 30 °C – +70 °C |
Best Temperature for stock (sensor): |
0 °C – +25 °C |
Condition pressure: |
600- 1400 hPa |
Condition humidity |
20 – 98 % r.h. |
Min Volume: |
1.0 L |
Battery: |
3.7V ,2200mAh rechargeable Li-ion battery |
Measurement time for one charge: |
More than 3000 times testing |
Dimension (LxWxH): |
201 mm x 57 mm x 64 mm |
Weight: |
200g include battery |
|
|
Other Parameter: |
|
Pre-heating time: |
about 5 seconds |
Zero time: |
about 3 seconds |
The previous measure is 0 mg/100ml: |
about 5 seconds |
The previous measure is 100 mg/100ml below: |
about 7 second |
Sensor type: |
Fast fuel cell |
|
|
Cross sensitivity: |
|
|
No |
Mouth pieces: |
|
|
Disposable, single way |
Units: |
|
|
‰BAC, mg/L BrAC, mg/100ml BAC, |
|
Add upon customer’s requests. |
Use interface: |
|
|
Menu Operation |
Input mode: |
|
|
3Main key+multi-function digital keyboard |
Power save mode: |
|
|
Automatic switch off after 2.5 minutes without use |
Display: |
|
|
color TFT display |
Memory: |
|
|
Up to 20,000 measured value with date and time |
Software Management: |
|
|
PC data manager, data transmit and Storage Option |
Calibration: |
|
|
Every 6 months or determined by local government |
|
|
|
|
Call Us :+91 960-381-5198
Mail :amruthtech@gmail.com
Reviews